Purpose: This study examined intraoperative imaging to monitor the alignment of the acetabular cup for total hip replacement arthroplasty (THRA) using C-arm fluoroscopy and to increase the accuracy of THRA prosthetic fixation.
Materials and Methods: From March 2004 to April 2007, 54 patients underwent THRA monitored by the imaging. 40° inclination and a 15° anteversion angle was considered to be ideal. The items compared included the alignment of the acetabular cup and the follow up rates of the hip dislocation.
Results: There was one case of hip dislocation. The mean inclination angle of the acetabular component was 39.9° with a standard deviation of 3.038. The mean anteversion angle of the acetabular component was 14.9° with a standard deviation of 1.398. The mean difference examination and variance in the anteversion angle showed a significant result . (P<0.05)
Conclusion: Intraoperative imaging increases the alignment accuracy of the acetabular cup and helps stabilize the prosthesis.
Keywords Total hip replacement arthroplasty, Acetabular cup, Alignment
J Korean Hip Soc 2007; 19(3): 155-160
Published online September 1, 2007 https://doi.org/10.5371/jkhs.2007.19.3.155
Copyright © The Korean Hip Society.
박상은∙김원유∙권오수∙김영율∙원호연
가톨릭대학교 의과대학 대전성모병원 정형외과학교실
Sang-Eun Park, M.D., Weon-Yoo Kim, M.D., Oh-Soo Kwon, M.D., Young-Yul Kim, M.D., Ho-Yeoun Won, M.D.
Department of Orthopaedic Surgery, Daejeon St. Mary’s Hospital, The Catholic University, Daejeon, Korea
Correspondence to:김원유
대전광역시 중구 대흥동 520-2 가톨릭대학교 의과대학 대전성모병원 정형외과
TEL: 82-42-220-9530
FAX: 82-42-221-0429
E-mail: weonkim@hotmail.com
* 본 논문은 2007년 가톨릭 중앙의료원의 연구 보조비 지원을 받아 이루어졌음.
Purpose: This study examined intraoperative imaging to monitor the alignment of the acetabular cup for total hip replacement arthroplasty (THRA) using C-arm fluoroscopy and to increase the accuracy of THRA prosthetic fixation.
Materials and Methods: From March 2004 to April 2007, 54 patients underwent THRA monitored by the imaging. 40° inclination and a 15° anteversion angle was considered to be ideal. The items compared included the alignment of the acetabular cup and the follow up rates of the hip dislocation.
Results: There was one case of hip dislocation. The mean inclination angle of the acetabular component was 39.9° with a standard deviation of 3.038. The mean anteversion angle of the acetabular component was 14.9° with a standard deviation of 1.398. The mean difference examination and variance in the anteversion angle showed a significant result . (P<0.05)
Conclusion: Intraoperative imaging increases the alignment accuracy of the acetabular cup and helps stabilize the prosthesis.
Keywords: Total hip replacement arthroplasty, Acetabular cup, Alignment